NOTICE
Water in the area of the electrical systems due to condensation
To avoid water collecting in the technical room due to condensation, the cold water pipes
must have suitable thermal insulation. Dripping water at the connections points must be
captured using appropriate collection containers.
If no in-house cold water system is available at the installation site, a local fixed,
decentralized cold water production system (inside or outdoors) can be included in the plans.
These systems are to be planned, installed, and maintained by local refrigeration technology
firms. Bruker BioSpin can help with planning and supplying these systems if necessary, but
not with installing and maintaining them.
Emergency cooling with drinking water
The supply of the MR system with cold water for operating the magnet is required continually
and without interruptions. In the event of a sustained breakdown of the mains cold water
supply, water from the
drinking water supply can be used temporarily
to cool the magnet
refrigerator. This can guarantee operation of the magnet over several days. For this reason,
Bruker BioSpin recommends installation of equivalent connection points for drinking water in
the technical room in the event of an emergency.
Summary of Contents for BioSpec 70/30 USR V2
Page 6: ...Contents vi Site Planning Information BioSpec 70 30 USR_2_V020...
Page 20: ...Safety 20 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 88: ...Moving or Dismantling the MR Instrument 88 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 98: ...Tables 98 Site Planning Information BioSpec 70 30 USR_2_V020...
Page 103: ...Site Planning Information BioSpec 70 30 USR_2_V020 103...