Adjustment of a Chamber on the D8
46
DOC-M88-EXX268_V1_12.2016
Figure 4.7: Inserting the temperature depending Z-positions
After saving the configuration, activating it and reconnecting, these values become active.
The Z-position can no longer be set independently, but it will now be set automatically (in
COMMANDER and WIZARD) according to its calibrated positions at the working tempera-
ture. The sample height is linearly interpolated for intermediate temperatures.
Note:
The Z-Position will only be updated if a new temperature outside the temperature
detection window (± 1 K for most chambers) is set.