Appendix / Service Details
100
DOC-M88-EXX268_V1_12.2016
8. Press once
Select
→ Turbo Yes (Backing Yes)
9. Press once
Menu
→ System ON Setup
10.Press once
Down
→ System Off Setup
11.Press once
Select
→ System Down
12.Press once
Down
→ Turbo Yes
13.Press once
Select
→ Backing No
14.Press once
Down
→ Backing No
15.Press once
Select
→ Backing No
16.Press once
Down
→ Backing Yes
17.Press once
Select
→ Backing Yes
18.Press once
Menu
→ System Off Setup
19.Press once
Menu
→ Turbo Off
20.Press
Cycle
twice for starting the system
For further information on programming the TIC controller see the Edwards manuals.
See also
2
Operation of the Edwards High Vacuum Stage [
10.2
Configuring a Chamber in DIFFRAC.MEASUREMENT
CENTER
10.2.1
HTK-series, TTK 600, XRK 900, and HPC 900 Chambers
The following procedure is suitable for the Anton Paar chambers HTK 2000N, HTK 16N,
HTK 1200N, TTK 600, XRK 900 and HPC 900. For this example, the configuring of an
HTK 1200N chamber is described. Other chambers can be configured by choosing file
names corresponding to the chamber name and its controller.