10 Appendix / Service Details
10.1
Vacuum Equipment
10.1.1
Programming of the Edwards High Vacuum Stage
The programming was done at Bruker already. If necessary use the following procedure to
program Edwards TIC controller, S/W issue: D39700640R. All settings will be reset by this
procedure!
Resetting the TIC to factory defaults:
1. Press once
Menu
2. Press 8 times
Down
→
Service Information
3. Press once
Select
4. Press 9 times
Down
→
Reset TIC Default
5. Press once
Select
→
Reset
Programming the TIC for automatic start and stop of the TMP and Backing pump:
Proceeding after resetting to factory defaults,
S/W issue: D39700640R
1. Press twice
Down
→ WRG New ID
2. Press once
Select
→ WRG and a pressure are indicated
3. Press twice
Up
→
Turbo Off
4. Press once
Select
→ Ramp up timer
5. Press 4 times
Down
→ System On Setup
6. Press twice
Select
→ Turbo No
7. Press once
Down
→ Turbo Yes