Installation and Operation of the CHC plus Humidity Chamber
72
DOC-M88-EXX268_V1_12.2016
´
This may take some minutes if the water bath has to he heated or cooled for a new
sample temperature
´
The example shows a heating chamber with Z-drive. It automatically corrects the
chamber height as function of temperature (height at 25 °C is 1 mm).
8. If this works successfully, you can start automatic measurements as described in your
Temperature Controller Manual
and in the manuals for DIFFRAC.SUITE [4] and
WIZARD [5].
9. Check box to wait for temperature and humidity reached before starting the scan.
Every time the Corio water bath was switched off, leave it switched off for at least 20
seconds and then switch on again!
Without temperature dependent height correction, peak positions might be wrong for
temperature >25 °C (alignment temperature).
Meaning of the CHC plus modes:
• Constant humidity: keeps the rel. humidity while changing the sample temperature.
This is a slow process because the water bath temperature has to be adapted.
• Set temperature and dry air flow
• Set temperature with rate
• Set temperature with rate and dry air flow
• Set temperature
• Set temperature and humidity
7.6
Alignment
All chambers discussed in this chapter can be aligned using the general method described
in chapter
Adjustment of a Chamber on the D8 [
.
Summary of Contents for Anton Paar
Page 6: ...Contents 6 DOC M88 EXX268_V1_12 2016 ...
Page 12: ...List of Figures 12 DOC M88 EXX268_V1_12 2016 ...
Page 26: ...General Remarks 26 DOC M88 EXX268_V1_12 2016 ...
Page 34: ...Vacuum Equipment 34 DOC M88 EXX268_V1_12 2016 ...
Page 135: ...DOC M88 EXX268_V1_12 2016 135 ...
Page 136: ...Bruker Corporation info baxs bruker com www bruker com Order No DOC M88 EXX268 ...