Installation and Operation of the CHC plus Humidity Chamber
62
DOC-M88-EXX268_V1_12.2016
Figure 7.5: Julabo Waterbath (Corio CD-201F)
Operation of the CHC plus requires:
• Electrical supply:
– TCU110 / CCU100 temperature controllers and MHG must be connected to the
mains board plugs X101.
– An adapter cable is used for connecting the MHG controller to the mains board
plugs X101.
– The Julabo HE-4 or Corio CD-201F heating circulator requires 230 V, 100 V or
110 V
• Cooling water (ca. 4 l, not destilled water) for Julabo inner circle supply
• External cooling water (ca. 1 l/min) for Julabo HE-4 heating circulator
• Compressed air (dry, oil-free, ca. 6 bar, 120 l/min) for sample cooling or LN2 for
cooling
• Transport gas: N
2
, dry air, He, CO
2
or Ar, 500 ml/min, ca. 2 bar
• Distilled water (ca. 1 l) for MHG water reservoir, for sample humidification