Brookfield Engineering Labs., Inc.
Page 56
Manual No. M13-167-A0415
Table
: Display all data points. Data will be shown in a scroll list where each page can hold a
maximum of 50 points. If the data set has more than 50 points, then additional pages
will be indicated at the bottom of the screen. Additional pages of data can be accessed
through the use of the Blue Arrows
Each data point includes: Viscosity, Torque, Speed, Temperature, Time, Shear Stress
(SS), Shear Rate (SR), Density, and Accuracy.
Test Used
: Display the test elements used to generate the data set. In this view, the Configure
Test button is available. Selecting Configure Test will program the DV3T to run
the same test utilized to collect the data set being viewed and present the Configure
Viscosity Test screen.
Notes
: Document any relevant information about the test or data. This information will be
stored with the data set once saved.
Compare Results
: Load two data sets and view side by side (see Figure IV-9).
Device Info
: Display basic information about the data file and the specific DV3T Rheometer
used to collect the data including: date and time the test started, completed and
was saved; if user accounts are active, which user saved the data; rheometer serial
number, firmware version, rheometer Torque range (LV, RV, HA, HB, other).
Graph
: Display the final data point and graph (see Figure IV-10). The graph axis (X and Y)
can be changed by selecting the blue drop down arrow in the appropriate data field.
Zoom and reset controls are indicated under the X-Axis.
Figure IV-10
Figure IV-9
Summary of Contents for DV3T
Page 2: ......