Brookfield Engineering Labs., Inc.
Page 10
Manual No. M13-167-A0415
7. Make sure that the AC power switch at the rear of the DV3T is in the OFF position.
Connect the power cord to the socket on the back panel of the instrument and plug it into
the appropriate AC line. For Cone/Plate Models, be sure that the toggle switch, used to
activate the electronic gap, is to the left position. (Left when facing the rheometer.)
The AC input voltage and frequency must be within the appropriate range as
shown on the nameplate of the rheometer. (See Section I.2.)
Note:
The DV3T must be earth grounded to ensure against electronic failure!!
8. Turn the power switch to the ON position and allow the rheometer to warm up for 10
minutes before performing autozero.
9. For Cone/Plate models, refer to Appendix A.
10. If appropriate, connect USB cable (DVP-202) to USB port for connection of DV3T
to PC or printer.
11. Review Read Me file. The Read Me file contains notes on the DV2T Firmware. This file
can be found on the CD that contains the PG Flash Software.
Temperature
Probe
USB B
USB A
On/Off
Switch
Ethernet
(Future Use)
Power Input
Temperature
Controller
Figure I-2