Fig. 12 Safely lifting the flaps
¾
Set the required cutting width by opening the
appropriate number of flaps
12
.
¾
Safely lift flaps
12
by pressing flap tabs
11
.
¾
Completely open flaps
12
until they engage in their
end positions (flap width = 8 mm).
Operating the Acculan® 3Ti Dermatome
Note
The low howling noise heard during startup of the
Acculan® 3Ti Dermatome is normal for this product!
¾
Press trigger
4
.
The dermatome runs at the continuously controlled
oscillation frequency.
Taking skin grafts
Before using the Acculan® 3Ti Dermatome for the first
time, the user should carry out some test cuts on a
specimen to make themselves familiar with the
operation of the device.
¾
Stretch flabby skin.
¾
With the sliding surface in level contact with the
skin, push forward dermatome
1
in a smooth
movement, applying gentle pressure. Be careful not
to cant dermatome
1
.
¾
For very thin grafts, use a forceps, from time to
time, to lay up the resected portion of the graft on
surface
13
.
¾
Switch off the motor and extract the fresh graft
from the dermatome.
- or -
¾
Lower the handle and cut off the skin graft with
dermatome
1
.
12
11
Summary of Contents for Acculan 3Ti
Page 2: ...23 26 25 21 22 22 20 24...
Page 3: ......
Page 209: ...207 2 7 6 24 10 9 3 1 20 3 24 9 7 6 20 1...
Page 210: ...Aesculap Power Systems Acculan 3Ti GA670 208 4 23 20 26 5 20 5 23 20 26 20...
Page 211: ...209 6 21 22 6 7 21 22...
Page 212: ...Aesculap Power Systems Acculan 3Ti GA670 210 8 9 1 22 21 21 7 23 8...
Page 213: ...211 9 1 21 25 9 1 25 23 25 25 23 23 10 1 4 4 5 OFF 10 4 1 4 5 ON 10 4 1 25 1 4 5 1...
Page 217: ...215 5 1 5 2 6 45 C VAH DGHM FDA CE pH 8 BH MAH GB228R...
Page 219: ...217 5 6 I II III T C F t I II III 1 Meliseptol HBV 50 1 IV 0 5 V...
Page 223: ...221 6 14 15 2012 6 B Braun Aesculap 14 Acculan 3Ti 15...