No.
Component
Article No.
④
Screwdriver
41796-27
NOTE: Any instrument adapter clamp can be attached to any array. If you exchange the array for
one of a different size during the procedure, recalibrate and verify the instrument before
continuing.
How to Attach the Array to the Instrument Adapter Clamp Softgrip
①
②
③
Figure 43
Step
1.
Select your
Instrument Adapter Clamp
and
Instrument Adapter Array
.
NOTE: The larger the instrument, the larger the array you should use.
2.
Insert the array stem
②
into the clamp
①
and align the teeth of both components.
3.
Tighten the screw knob
③
by hand or by using the
Screwdriver
, ensuring that the array
is securely attached to the clamp.
Instrument Adapter Clamp Softgrip (Size L, XL)
84
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking