Warning
If using marker spheres on, or in the vicinity of, oscillating or vibrating instruments, the
marker spheres must be checked at regular intervals to make sure that they remain
securely attached.
Warning
Set up the OR considering permanent visibility of all navigated equipment.
The cameras should have an unobstructed view of the reflective marker spheres on the
tracking array. The marker spheres shall not be masked. Otherwise, navigation is not
possible or inaccuracies may occur during navigation. Plan the OR setup prior to surgery.
The camera must have an unobstructed view of the reflective marker spheres of the
instrument.
NOTE: Ensure that marker spheres are visible to camera at all times during use.
Array Positioning
The system recognizes an instrument by the geometrical arrangement of the marker spheres on
its tracking array. For the system to properly identify an instrument, the camera must have a clear
view of its marker spheres at all times.
For the best recognition of the array, position the array perpendicular to the plane through the
centers of all marker spheres.
Possible Complications
Use of some Brainlab products requires an additional incision to be made. Therefore, the following
complications may occur: infection, local pain, bleeding, lesion of blood vessels or nerves, bone
fracture or thrombosis.
Creutzfeldt-Jakob Contamination
Warning
Do not use Brainlab instrumentation on patients suspected of having Creutzfeldt-Jakob
disease (CJD or vCJD).
Sterilization
If a sterile and reusable instrument or any of its components is inadvertently removed from the
sterile field, or comes into contact with an unsterile object during unpacking or clinical use, it must
be sterilized again.
Do not resterilize or reuse disposable instruments, as this could affect navigation accuracy, posing
a risk to the patient. They must be disposed of after each use.
Warning
Unless otherwise indicated, instruments must be sterilized before use. Detailed information
about reprocessing procedures, parameters and cleaning detergents are available in your
Cleaning, Disinfection and Sterilization Guide. Reprocess reusable devices according to
the descriptions in the Cleaning, Disinfection and Sterilization Guide.
Instrument Contact to Brain Tissue and Spinal Fluid
Only the following instruments are intended for direct contact with brain tissue and spinal fluid:
•
Disposable Biopsy Needle 1.8 mm/235 mm (5 PCS)
(41779)
•
Disposable Stylet (10 PCS, Pre-Calibrated)
(55797)
•
Disposable Pre-calibrated Suction Charr. 8 (10 PCS)
(52184)
•
Disposable Pre-calibrated Suction Charr. 12 (10 PCS)
(52185)
Instrument Overview
14
Instrument User Guide Rev. 2.6 Cranial/ENT Optical Tracking
Summary of Contents for CRANIAL/ENT OPTICAL TRACKING
Page 2: ......
Page 28: ...Instrument Calibration Matrix 28 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Page 52: ...Electromagnetic Compatibility 52 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Page 94: ...Ultrasound Calibration 94 Instrument User Guide Rev 2 6 Cranial ENT Optical Tracking ...
Page 110: ......
Page 111: ......