Operation
4.1 Before use
1.
Switch on the HF device.
2.
Remove the product from the sterile packaging.
3.
Connect the HF cable to a bipolar socket of the HF device
4.
The output power and mode is automatically set on the
BOWA ARC generator with
ЭСО PХuР’n CuЭ COMŻORT
instrument detection.
5.
Perform a thorough visual inspection and functional test
before using the ligation instrument.
4.2 Inspection
1.
Check if the rotation wheel can be turned easily until the
stop is reached.
2.
Check the jaws for proper opening and closing using the
lever.
3.
Lock the handle by pressing the lever down.
4.
Pull the trigger to check for ease of motion.
5.
Check the finger switch for proper activation.
The activation signal sounds while the finger switch is
being pushed.
WARNING
Risk of patient injury!
Use only intact products.
Risk of patient injuries from the combustion or
explosion of flammable liquids and gases!
Avoid contact with flammable gases and liquids, such
as skin cleaners, disinfectants and anaesthetic gases
Summary of Contents for ERGO 310D
Page 1: ...Gebrauchsanweisung DE...
Page 15: ...Instructions for use EN...
Page 29: ...Mode d emploi FR...
Page 43: ...Instrucciones de uso ES...
Page 57: ...Istruzioni d uso IT...
Page 71: ...Kullanma talimat TR...
Page 85: ...Ins ruk j o s u PL...
Page 99: ...RU...
Page 100: ...MN031 600 S0 RU ERGO 310D 2...
Page 101: ...MN031 600 S0 RU ERGO 310D 3 1 1 1 BO A ARC BOWA ARC 400 350 7 1 2...
Page 102: ...MN031 600 S0 RU ERGO 310D 4 7 7 7...
Page 103: ...MN031 600 S0 RU ERGO 310D 5 2 2 1 8 13...
Page 104: ...MN031 600 S0 RU ERGO 310D 6 2 2 BO A 2 3...
Page 105: ...MN031 600 S0 RU ERGO 310D 7 2 4 BO A ARC 2 5...
Page 106: ...MN031 600 S0 RU ERGO 310D 8 3 ERGO 310D 220...
Page 107: ...MN031 600 S0 RU ERGO 310D 9 4 4 1 1 2 3 4 BO A ARC 5 4 2 1 2 3 4 5...
Page 108: ...MN031 600 S0 RU ERGO 310D 10 4 3 1 2 3 4...
Page 109: ...MN031 600 S0 RU ERGO 310D 11 5 6 7 8 9 10 1...
Page 110: ...MN031 600 S0 RU ERGO 310D 12 4 4 1 2 4 5 5 6 1 2...