10 Technical specifications
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
89
10. Technical specifications
10.1.
ARC 400 technical data
Insulation type / Classification
EMC
IEC 60601-1-2
Level of protection provided by the housing
IP 21
Protection class according to EN 60601-1
I
Application component type according to
EN 60601-1
CF
Standards compliance
IEC 60601-1: 2005,
IEC 60601-1-2: 2007,
IEC 60601-2-2: 2009,
ISO 14971: 2007,
ISO 13485: 2003 +
Cor.1_2009
Classification according to EC Directive
93/42/EEC
IIb
Power connection
Power consumption in standby mode
40 W / 85 VA
Line frequency
50 / 60 Hz
Maximum power consumption with 300 W HF
output power
550 W / 975 VA
Terminal for potential equalization
Yes
Voltage range 220 V – 240 V
Input voltage range
198 V to 264 V
Current consumption in standby mode
110 mA
Current consumption at maximum HF output
4 A
Line fuses
2 x 5 A slow-blow
Voltage range 100 V - 115 V
Input voltage range
90 V to 130 V
Current consumption in standby mode
220 mA
Current consumption at maximum HF output
8 A
Line fuses
2 x 10 A slow-blow
Dimensions and weight
External dimensions: width x height x depth (mm) 430 x 180 x 475
Weight
approx. 12.5 kg