10 Technical specifications
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
93
Mode icon
Description
CCS
ARC
CONTROL
Form of HF
voltage
Max. power output
Peak
voltage
Default values
Effect
Power range
Effect
Max. Watt
Modes Monopolar Cutting
GastroCut
Loop fast
Yes
Yes
sinusoidal
alternating
Coag, Cut
and break
phases
1
2
3
4
5
6
7
8
9
-
750 Vp
5
---
GastroCut
Knife slow
Yes
Yes
sinusoidal
alternating
Coag and Cut
phases
1
2
3
4
5
6
7
8
9
-
650 Vp
5
---
GastroCut
Knife medium
Yes
Yes
sinusoidal
alternating
Coag and Cut
phases
1
2
3
4
5
6
7
8
9
-
650 Vp
5
---
GastroCut
Knife fast
Yes
Yes
sinusoidal
alternating
Coag and Cut
phases
1
2
3
4
5
6
7
8
9
-
650 Vp
5
---
Modes Monopolar Coagulation
Moderate
sinusodial
constant
1
2
3
1 W - 120 W
250 Vp
2
60
Forced non
cutting
pulsed
modulated
-
1 W - 80 W
3.5 kVp
---
50
Forced mixed
sinusoidal
modulated
1
2
3
1 W - 120 W
2.3 kVp
2.5 kVp
2.8 kVp
2
60
Forced cutting
sinusoidal
modulated
1
2
3
4
1 W - 250 W
1.5 kVp
1.5 kVp
1.3 kVp
1.3 kVp
2
80
Spray
pulsed
modulated
1
2
3
4
1 W - 120 W
3.0 kVp
3.8 kVp
4.6 kVp
5.0 kVp
2
80
Argon open
pulsed
modulated
-
1 W - 120 W
4.6 kVp
---
80
Argon flexible
pulsed
modulated
-
1 W - 120 W
4.4 kVp
---
40