5 Operation
56
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
5.7.8. Argon flex. pulse
This mode is used for argon-assisted electrosurgery in conjunction with the ARC
PLUS accessory device.
This is the current type Spray.
For argon-assisted coagulation, flexible probes are used in combination with
endoscopes.
The puls frequency changes with the effect setting. The higher the effect level, the
faster the pulse sequence.
Application areas
Gastroenterology, homogeneous surface coagulation
Suitable instruments
Flexible argon probes
5.7.9. Resection
This mode is used for monopolar hemostasis in gynecology and urology.
Use non-conductive irrigation fluids.
Application areas
Hysteroscopy, transurethral prostate resection (TUR-P), surgical treatment of
bladder tumors (TUR-B), vaporization of prostate tissue (TUR-VAP).
Suitable instruments
Resectoscope (monopolar)
Resection sling
Rollerblade electrode