![Bowa ARC 400 Operating Manual Download Page 50](http://html1.mh-extra.com/html/bowa/arc-400/arc-400_operating-manual_2797070050.webp)
5 Operation
50
Operating Manual ARC 400
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
5.6.9. GastroCut Loop medium
This mode is used in gastroenterology. Polypectomy snares are used for cutting
and coagulation. ARC control generates the cutting effect with simultaneously
minimized output power. This mode consists of a series of cutting current pulses
followed by a coagulation phase. With an accelerated pulse rate of 1.8 cutting
pulses per second, this mode is suitable for experienced users.
Application areas
Polyp removal using polypectomy snares and flexible endoscopy, with accelerated
pulse rate for experienced users.
Suitable instruments
Polypectomy snares
5.6.10. GastroCut Loop fast
This mode is used in gastroenterology. Polypectomy snares are used for cutting
and coagulation. ARC control generates the cutting effect with simultaneously
minimized output power. This mode consists of a series of cutting current pulses
followed by a coagulation phase. With an accelerated fast pulse rate of 3.2 - 5
cutting pulses per second, this mode is suitable for advanced users.
Application areas
Polyp removal using polypectomy snares and flexible endoscopy, with accelerated
fast pulse rate for advanced users.
Suitable instruments
Polypectomy snares