![Bowa ARC 400 Operating Manual Download Page 47](http://html1.mh-extra.com/html/bowa/arc-400/arc-400_operating-manual_2797070047.webp)
5 Operation
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
47
5.6.3. Dry
This mode is used for monopolar dry cutting. A large, controlled arc is generated,
which allows significantly deeper coagulation to be obtained.
Application areas
Cardiac surgery and blood coagulation in retracting blood vessels in the sternum
region.
Suitable instruments
Knife electrodes
5.6.4. Argon
This mode is used to perform open surgical interventions in combination with the
ARC PLUS companion device for argon-assisted cutting. With suitable instruments
connected, argon-assisted cutting can be performed using rigid electrodes.
Application areas
Visceral surgery
Suitable instruments
Rigid argon electrodes
Argon handpiece