![Bowa ARC 400 Operating Manual Download Page 51](http://html1.mh-extra.com/html/bowa/arc-400/arc-400_operating-manual_2797070051.webp)
5 Operation
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
51
5.6.11. GastroCut Knife slow
This mode is used in gastroenterology. Instruments for papillotomy and
endoscopic resections are used for cutting and coagulation. ARC control
generates the cutting effect with simultaneously minimized output power. This
mode consists of a pulse sequence of cutting current and coagulation phase. With
a relatively slow pulse rate of 1.3 cutting pulse per second, this mode is suitable for
especially cautious work.
Application areas
Papilla incision using a papillotome and flexible endoscopy, resection with needle
knives; slow pulse rate for cautious work.
Suitable instruments
Papillotome
Needle knives
5.6.12. GastroCut Knife medium
This mode is used in gastroenterology. Instruments for papillotomy and
endoscopic resections are used for cutting and coagulation. ARC control
generates the cutting effect with simultaneously minimized output power. This
mode consists of a pulse sequence of cutting current and coagulation phase. With
an accelerated pulse rate of 1.7 cutting pulses per second, this mode is suitable for
experienced users.
Application areas
Papilla incision using a papillotome and flexible endoscopy, resection with needle
knives; accelerated pulse rate for experienced users.
Suitable instruments
Papillotome
Needle knives