5 Operation
900-400_IFU_V1.0.0.6.17.9_20140-S2-20130604-EN
Operating Manual ARC 400
41
Save current program on Dr. Dongle:
The program currently loaded in ARC 400 can be saved on Dr. Dongle:
1. In the overview, select a program storage space to be overwritten.
2. Press “save” to save the currently loaded program at the selected space.
3. A keyboard appears which can be used to enter a name for the program.
The program is now saved on Dr. Dongle.
4.
Press the "back" button to return to the main screen.
Dr. Dongle can now be removed
In order to save another program, it is necessary to either carry out
a program change or to remove and re-insert Dr. Dongle.
The aforementioned steps 1-4 can now be repeated.
Load a program from Dr. Dongle:
A program saved on Dr. Dongle can be loaded onto any ARC 400:
1. Select a program saved on Dr. Dongle from the overview by touching the
program name.
2. Press "load" to load the selected program.
- or-
Press "back" to return to the main screen without carrying out
changes to the selection.
The selected program is now active in the main screen of ARC 400.
Dr. Dongle can be removed
- or-
Save the loaded program in the program list, see "Save program" dialog,
page 72.
5.4.8. Plug’n Cut COMFORT
The automatic instrument identification Plug’n Cut COMFORT recognises the
connected BOWA COMFORT instrument and selects the default parameters
automatically.
1. Insert the COMFORT instrument into a socket of ARC 400.
The instrument data is read