5.9.3. Micro forceps
This mode is used for arcless contact coagulation with micro forceps. It enables
extremely fine control of power output down to 0.1 W and precise work for tightly
restricted bipolar contact coagulation.
Application areas
Bipolar coagulation in pediatric surgery, neurosurgery, plastic surgery, etc.
Suitable instruments
Bipolar forceps
Micro forceps
5.9.4. Forceps forced
This mode is used for forced coagulation with forceps.
Application areas
Fast bipolar coagulation
Suitable instruments