1 689 989 485 | 2020-08-01
Robert Bosch GmbH
5.3
Attaching the calibration board in
vertical position D
The mounting adapters are used for attaching all
calibration boards CTA 3xx-x to the
DAS 3000 S20.
---
1. Mount the mounting adapter
(1)
at vertical position
D.
2. Make sure the mounting adapter correctly engages
with the mounting pins
(3)
by the mounting guides
(2)
.
3. Use the knurled screw
(4)
to hand-tighten the
mounting adapter.
1
22
33
4
33
4
4712009-04_shd
(1) Mounting adapter
(2) Mounting guides
(3) Mounting pin
(4) Knurled screw
4. Position calibration board retainer Z (1) on the cen‐
ter vertical bar below vertical position D.
CAUTION – Pinch point hazard when positioning the
calibration board.
5. Do not place fingers or hands between the mounting
adapter
(2)
and the bottom edge
(3)
of the calibra‐
tion board.
6. Place the bottom edge of the calibration board
(3)
in
the mounting adapter
(2)
on the positioning unit
(7)
.
7. Push calibration board retainer Y
(5)
onto the top
edge
(4)
of the calibration board in such a way that
the recess in calibration board retainer Y secures the
calibration board.
8. Use the knurled screw
(6)
to hand-tighten calibra‐
tion board retainer Y.
The calibration board is now secured and fixed in
place.
4712005-27_shd
1
2
3
4
5
6
7
(1) Calibration board retainer Z
(2) Mounting adapter
(3) Bottom edge of calibration board
(4) Top edge of calibration board
(5) Calibration board retainer Y
(6) Knurled screw on calibration board retainer Y
(7) Positioning unit
9. Make sure that the arrow and marking
(2)
on the
bottom edge of the calibration board align with the
mounting adapters
(1)
.
The calibration board is centered relative to the
DAS 3000 S20.
1
22
1
22
4715002-28_shd
(1) Mounting adapter
(2) Arrow and mark on calibration board for centering
---
5.4
Attaching the indicator bar to the
frame
CAUTION
Magnetic fields from strong magnets on the in‐
dicator bar. Minor injury to people with a pace‐
maker or implanted defibrillator.
People with a pacemaker or implanted defibril‐
lator must maintain a sufficient distance from
the DAS 3000 S20, especially from the indicator
bar.
---
No calibration board is attached at vertical positions
A, B, C or D.
Calibration board CTA 300-1 is in the parked posi‐
tion.
DAS 3000 S20 | 29 | en