10-2 Analog output specification and connection method (DAV, DAI)
For DAV(DAI) output, the measurement value is converted into D/A and output as analog voltage (current).
Output range include 0~5V, 0~10V, 0 ~ ±5V, 0 ~ ±10V, 0~20mA, and 4~20mA.
Output scaling and minute calibration is set from function.
Frequency trait is based on display conversion speed setting of function.
10-2-1 Specification
Range of rated output
–10 V ~ +10 V
4 ~ 20mA
Range of max output :
–12.0 V ~ +12.0 V
0 ~ 24mA
D/A conversion speed :
1000
times/sec
1000 times/sec
Valid resolution :
1/10000
1/10000
Zero drift :
±1 mV/°C typ.
±0.5uA/°C typ.
Gain drift :
±100ppm/°C typ.
±100ppm/°C typ.
Non-linearity :
Within ±0.05%F.S.
Within ±0.05%F.S.
Load resistance ::
5 k
Ω or higher
500
Ω or higher
10-2-2 Connection method
(DAV)
Terminal No.
Contents
1. VOLT +
Voltage output +
2. VOLT -
Voltage output -
10-2-3 Connection method
(DAI)
Terminal No.
Contents
3. CURR -
Current output -
4. CURR +
Current output +
Analog output is to convert measurement value into D/A for output and display conversion speed shall be
changed from function setting in case of wanting 1000 times/sec output.
It is output based on display conversion speed of function setting.
In case of
DAI analog output(4~20mA) (-) display value is not output.
Output at function mode entry maintains previous value before entry and there is no output change.
Voltage and current output cannot be used at once.
Either voltage or current mode shall be selected from function setting for use. (it is different output terminal).