High Precision DIGITAL INDICATOR
M
High Precision DIGITAL INDICATOR
M
High Precision DIGITAL INDICATOR
M
High Precision DIGITAL INDICATOR
M
High Precision DIGITAL INDICATOR
M
High Precision DIGITAL INDICATOR
M
High Precision DIGITAL INDICATOR
Select Unload among Load and
Unload with the use of No.2
and No.3 key and push No.4
key when there is nothing on
load cell.
Conduct zero calibration.
If the zero setting has already
been made and the weight is
placed on the weighbridge,
you can proceed by selecting
load without unloading.
It is changed to dead weight
value mode when pushing
No. 4 key after load display.
Input dead weight value with
the use of No. 2 and No,.3
key.
ex) 5.000kg
Push No.4 key after exerting
actual load to load cell.
Actual load calibration has
been completed.
Err0 is played and it is turned to LoAd
status when load is small or direction
of load is applied toward (-) direction.
Check the dead weight value indicated to display.
Check the zero point return status after lowering the dead weight and
Step2 shall be executed repeatedly when there is abnormality.
Warning
Display speed shall be set as slow as possible in function mode during calibration in order to
display stable value.
Although resolution setting range is 1/30000 or lower, display is still conducted when it exceeds 30000.
Conduct regular check on measurement and conduct calibration upon necessity.
Calibration under instable environment may cause the measurement error.
It is recommended that the use of dead weight with over 2/3 of max capacity is recommended
in order to reduce the error.