6-4 Linearization Calibration Mode
6-4-1 Calibration Method
A little measurement error may occur in between the max capacity due to trait of measurement
unit even upon the completion of zero and span calibration. Linearization calibration mode is a
type of calibration mode which conducts non-linear calibration that reduces the measurement
error by conducting calibration with max 4 points excluding the zero with the use of actual load.
ZERO
0
Weight Value
Di
sp
la
y
Error
ZERO
Before linear compensation
After linear compensation
0
Weight Value
Di
sp
la
y
Calibration point 1 (Load1)
Calibration point 2 (Load2)
Calibration point 3 (Load3)
SPAN
SPAN
Step 1. Decimal point setting and min gradation setting
Enter into function mode by
pushing No.2 key while
pushing No.1 key.
It is changed to decimal point
mode when pushing No. 4 key.
After ‘
dP
’ display,
Push No.,4 key after changing
decimal point by pushing
either No.2 or No.3 key.
After ‘
dIv
’ display,
Input is completed when you
push No.4 key after changing
min unit by pushing either
No.2 or No.3 key.
After ‘
rAtE
’ display,
It is changed to measurement
mode when pushing No.1 key..
Decimal point setting and min.
unit setting have been
completed.