6-5-3 Zero Tracking
Automatic update of zero point shall be conducted by detecting the movement of zero point with the use
of zero tracking function (setting shall be conducted at function).
Zero tracking shall only be conducted when width and time function of zero tracking are already set.
Zero tracking is not conducted when zero point is within zero calibration range.
Zero Calibration
Width of Zero Tracking
nonE None
Width of zero tracking shall be set.
0.000 Setting Range
0 ~ 999
Time of Zero Tracking
00 Setting Range
0 ~ 99
Zero tracking operation time shall be set.
00: Operation within 0.1 sec.
01: 0.1 sec
~
99: 9.9 sec
Warning: When ‘None’ is selected at zero
tracking width setting, zero tracking time
setting mode shall not be displayed.
Display 0
Tracking width
Start
Tracking time
Zero tracking does not operate when the maximum value of the zero adjustment range exceeds 999.
Zero tracking is not operated when offset value is set at function.
When there is a dramatic change in load by vibration even within the range of zero tracking width, zero
tracking setting times is doubled thus there may be a delay in conduct of automatic zero calibration.