11.2
When Error Message is Displayed on LCD
11.2.1 Error
Messages
When error is detected, "Error n
˚
**/**" and the error message are displayed on
the lower line of the LCD alternately. The first number of "**/**" is the order of
detected errors that have occured. Thus, a "1/3" would denote the fastest error
of 3 errors. Press the "UP" or "DOWN" switch to scroll display the contents of
the next or previous error message. (The "UP" or "DOWN" switch can also be
input in Remote Mode.)
Example)
Press
"DOWN" Switch
BRAKE : ***00rpm
Error n
°
1/3
BRAKE : ***00rpm
Error n
°
2/3
Press "UP" Switch
BRAKE : ***00rpm
Disturbance Y_H 1/3
BRAKE : ***00rpm
Mains Failure 1/3
Displayed alternately
Press "UP" Switch
Displayed alternately
T
ro
ub
le
sh
oo
tin
g
◇
The upper message on the LCD shows the current STP pump
operation state.
◇
When the message on the LCD is in Conformation mode or
Parameter Set mode, press the "SELECT" switch several times to
change it to the status display mode.
11-2