![BOC Edwards Seiko Seiki STPH1303C Instruction Manual Download Page 75](http://html1.mh-extra.com/html/boc-edwards/seiko-seiki-stph1303c/seiko-seiki-stph1303c_instruction-manual_2781984075.webp)
STP-H803/H1303 Series Instruction Manual
9.2 Confirmation
Mode
Confirmation Mode is used to check the status of the STP pump and STP control
unit. The following items can be checked in Confirmation Mode:
(1) Version information
(2) Individual information (serial number, total hours of running, and
number of starts)
(3) Current settings (rated speed, TMS temperature setting, and acutual
pump temperature)
(4) Current function settings (remote mode, TMS function, inhibit command,
and emergency vent. valve)
(5) Error record (lists up to 10 of the most recent errors)
O
pe
ra
tio
n
◇
The emergency vent. valve is not used with this series of STP pump.
<Confirmation Method> (See Figure 9.2)
1) Press the "SELECT" switch to enter Confirmation Mode.
The available menu groups (items (1) through (5) above) are displayed in
order.
2) Press the "UP" switch to display the next item. Press the "DOWN" switch to
display the previous item.
3) Press the "SELECT" switch to display the next menu group.
4) Press the "UP" and "DOWN" switch to confirm the status of the items of the
currently selected menu group.
5) Press the "SELECT" switch to return to Operation Mode when viewing the
Error Record. Or, if no switch is pressed for approx. 1 minute, the display
will automatically revert to Operation Mode.
9-4