13.2
Specifications for the STP Control Unit
Table 13.2 Specifications for the STP Control Unit (1/2)
Specification
Item Specifications
Input voltage
V AC
200 to 240
±
10%
Without the TMS unit:
VA
Max. 850
Input power
With the TMS unit:
VA
Max. 1,200
Input frequency
Hz
50/60
±
2
Leakage current value
mA
Max. 3.5
Input phase
Single phase
Rated current
A
15
Main breaker spec.
AIC: Ampere Interrupting
Capacity A
1000 (AC240V, 50/60Hz)
Motor drive system
3-phase DC brushless motor driver
Output voltage under rated operation
V AC
79
Output frequency under rated operation
Hz
542
Ambient temperature range
°C
0 to 40
Storage temperature range
°C
- 25 to + 55
9
Dimensions
mm
W210
×
H132.5
×
D401.6
TMS control unit
Internal
Operation switches
•
START switch
(flat panel switch, green)
•
STOP switch
(flat panel switch, dark gray)
•
RESET switch
(flat panel switch, gray)
•
SELECT switch
(flat panel switch, gray)
•
UP switch
(flat panel switch, gray)
•
DOWN switch
(flat panel switch, gray)
•
ENTER switch
(flat panel switch, gray)
•
MANUAL/REMOTE changeover switch
(slide switch, black)
Panel indication lamps
•
ACCEL lamp
(green LED)
•
NORMAL lamp
(green LED)
•
BRAKE lamp
(green LED)
•
TEMP CTRL lamp
(green LED)
•
POWER lamp
(green LED)
•
FAILURE lamp
(red LED)
•
REMOTE lamp
(green LED)
Panel display
•
LCD panel (20
×
2 characters)
*1
Mass is a value of state that the only standard accessory was installed (except the optional accessory).