Model BD306 Dual-Input Pump Controller Instruction Manual
Error Message (
Error
)
An error message indicates that the calibration or scaling process was not successful.
After the error message is displayed, the meter reverts to the input 2 during calibration or
scaling and to input 1 during internal calibration, allowing the appropriate input signal to be
applied or programmed.
The error message might be caused by any of the following conditions:
1.
Input signal is not connected to the proper terminals or it is connected backwards.
2.
Wrong signal selection in
Setup menu.
3. Minimum input span requirements not maintained.
4. Input 1 signal inadvertently applied to calibrate input 2.
Minimum Input Span
The minimum input span is the minimum difference between input 1 and input 2 signals
required to complete the calibration or scaling of the meter.
Input Range
Input 1 & Input 2 Span
4-20 mA
0.15 mA
±10 VDC
0.10 VDC
7.7.4.
Calibrating the Meter with External Source (
CAL
)
To scale the meter without a signal source
refer to Scaling
the Meter without a Signal
Source
, page 38.
Warm up the meter for at least 15 minutes
before performing the calibration to ensure
specified accuracy.