Title: Final Test 8500
Document No: 129TP000
Sheet 13 of 17
BLEASE QUALITY DOCUMENT
BQD 083.V2
25/5/94
Checked:-________
5.24 Dealer specific information
The following options can be set:
Absorber
Present (default) / Not present
Language
English (default), German, French, Spanish, Turkish, Italian
O.E.M
Blease (default), A.M.S, Acoma, Finesa, User select
The Absorber switch setting can be found under setup enter the password “TECHY”
The Language and O.E.M settings can also be found under setup enter the password “LINGO”
If anything other than the default options are required then these should be set
All options should be recorded on the calibration record even if they are default.
5.25 Test Completion
Restore ventilator controls to default. Ensure mode on/off switch is in the OFF.
Check that all the noted readings are within the tolerances as specified in section 4.
Sign off the 'Checklist and Calibration Record' and the customer printout. Ensure both have the
ventilator serial number, date and signature of final tester, etc.
If the original sticker over the hole in the calibration port on the rear of the ventilator was
removed, replace with a new ‘calibration void’ sticker.
Attach the required notification to the ventilator which states that the unit has passed its final
test.
Calibration Gas
As specified by model number
Gas Convert
Norm
BAV Software Version
8.60
Front Panel Software Version
8.20
Ventilator Model No:
Manufacture Date
(part of serial no.)
Calibration Date:
All options on second sheet are at default.
5.26 Printer Output
Connect the serial printer to the recorder output.
With no pneumatic connections attached, turn the ventilator ON (adult mode) and check that
the printer starts operating after a few seconds. Note the output software version on the
checklist and calibration record and any other details required in the record header.