Title: 8500 Adjustment & Calibration
Document No: 129AC000
Sheet 9 of 12
BLEASE QUALITY DOCUMENT
BQD 083.V2
25/5/94
adjust flow for a peak pressure of 80 cmH
2
O. Wind R99 on the BAV board fully clockwise.
Adjust R98 on the BAV board until pressure limiting is observed. This can be checked by
observing that the can of Q1 goes high to low each time the limit is reached.
Once triggering has been achieved, decrease delivered volume to reduce the peak pressure to
find the exact trigger point. The trigger point should be such that it will not have any effect on
the ventilator functions. It should be set to greater than 80cmH
2
O.
4.13 Fresh Gas Flow Sensor Calibration
Ensure that the unit is standby.
Run the Blease Medical Test I/O Diagnostics, connect the ventilator serial cable between the
calibration port connector on the back panel of the 8500 & a PC. Run HyperTerminal on the PC
using the stored 8500 settings file. Press Ctrl+E to connection to 8500 then press up arrow to
enter calibration page.
Ensuring that there is no fresh gas flowing, highlight the Fresh gas flow zero field and press
spacebar to set value.
Set 10 LPM fresh gas flow, highlight Fresh gas flow gain (10 L/M) field and press spacebar to
set value. Press Ctrl+U to save values.
Press Ctrl+E to exit calibration page.
Ensure that the unit is standby.
Press the memory button, then select configuration.
Select setup and enter password ‘TECHY’
Select fresh gas.
Ensure there is no flow through fresh gas flow sensor.
Press trakwheel to set 0 LPM
Set each flow rate as requested, press trakwheel to set.
Return to main screen.
4.14 Saving Data and Printer Check
Check all printer output parameters are set up and there are no warning messages. Label
ventilator as calibrated.
Enter calibration date, on diagnostics page 2.
Enter the serial number & and the model number of diagnostics page 2.
Press Ctrl + U to save.
Press Ctrl + E to complete software calibration.
Run the Blease Medical Test I/O Diagnostics, connect the ventilator serial cable between the
calibration port connector on the back panel of the 8500 & a PC. Run HyperTerminal on the PC
using the stored 8500 settings file.
Choose printer output from the transfer menu. A print out of all parameters just set up is
automatically obtained at switch on. Deselect printer output on the transfer menu, the stored
information will now be printed.