Operations & Maintenance Manual: Product Specifications
Table 24. Accuracy Specifications
SPECIFICATION
DESCRIPTION
Bladder volume accuracy
± (15% + 15 mL)
Bladder volume range
0–999 mL (0–200 mL in small child scan mode)
The following example shows how the accuracy ranges shown in
Table 24
could affect reported volume
measurements.
If the measurement scanned is 160 mL, with the accuracy range of ± (15% + 15 mL), the volume is calculated
as follows:
160 mL × 15% = 24 mL
22.5 mL + 15 mL = 39 mL
150 mL ± 39 mL =
111–189 mL
The accuracy specifications assume the instrument is being used according to the instructions provided by
Verathon
®
while scanning a tissue‑equivalent phantom.
Table 25. Operating & Storage Conditions
SPECIFICATION
DESCRIPTION
Operating Conditions
Ambient temperature range
10–40ºC (50–104ºF)
Atmospheric pressure range
700–1060 hPa
Relative humidity
30–75% non‑condensing
Ingress protection (IP) against
water
Console
IPX0
Probe
IPX1
Storage Conditions
Ambient temperature range
‑20–60ºC (‑4–140ºF)
Atmospheric pressure range
500–1060 hPa
Relative humidity
20–95% non‑condensing