65
Operations & Maintenance Manual: Product Specifications
PRODUCT SPECIFICATIONS
COMPONENT SPECIFICATIONS
CONSOLE & PROBE SPECIFICATIONS
Table 21. General Specifications
ITEM
SPECIFICATION
Input
Lithium‑ion battery
Classification
Internally powered, Type BF
Expected product life
5 years
Weight
2.36 kg (5.2 lbs), with battery
Display
13.36 x 10.13 cm (5.26 x 3.99 in) (640 x 480 pixels, 120 dpi)
Integrated printer
Thermal printer
Table 22. Ultrasound Acoustic Output Parameters (FDA Format)
Values in this table are the maximum readings obtained from three test results.
ACOUSTIC OUTPUT
MI
I
SPTA.3
(mW/cm
2
)
I
SPPA.3
(W/cm
2
)
Global Maximum
Value
0.519*
0.632
9.35
Associated
Acoustic
Parameter
p
r.3
(MPa)
0.684
W
o
(mW)
1.55
1.44
f
c
(MHz)
1.74
1.74, 2.63
†
1.74
z
sp
(cm)
1.90
1.90
Beam
dimensions
x
‑6
(cm)
0.321
y
‑6
(cm)
0.334
PD
(µsec)
2.93
2.93
PRF
(Hz)
400
400
EDS
Az. (cm)
7.40, 7.38
†
Ele. (cm)
7.40, 7.38
†
TIS/TIB/TIC range
0.0‑1.0*
* Both MI and TI values are below 1.0.