Operations & Maintenance Manual: Troubleshooting
TROUBLESHOOTING PROCEDURES
PROCEdURE 1. TROUBLESHOOT SCAN POINT CONNECTION
Complete this procedure if the console cannot connect to Scan Point
®
.
1. Ensure the
Scan Point
button
was pressed and the console is searching for a connection.
2. In Scan Point, try connecting again by clicking the
Find New
button. Repeat this step up to 3 times.
If the console does not connect, continue to the next step.
3. Turn the console off, turn the console on, and then press the
Scan Point
button
. On the PC, in
Scan Point, click
Find New
.
If the console does not connect, continue to the next step.
4. While the instrument is attempting to connect to Scan Point, remove the battery.
5. Reinsert the battery, allow the instrument to power on, and then press the
Scan Point
button
.
6. On the PC, click
Find New
.
If the console does not connect, contact Verathon
®
Customer Care.