Revision 15.00 / LB-0028
–
Cerebus Instructions for Use
5
Packing Contents
The Cerebus System is shipped with components listed below. The exact type and quantity of the
components can vary depending on the Cerebus model, line voltage (110v or 220v), and total input
channel count.
•
Neural Signal Processor (NSP) and rubber feet
•
Rack mounting ears and screws
•
For Cerebus System only: Front End Amplifier (FEA) and isolated Amplifier Power Supply
(APS).
•
For Digital Cerebus System only: Digital Hub 128 and isolated power supply.
•
Alligator clip to touch proof ground cable
•
Power cables and connectors
•
Crossover Ethernet cable
•
Cerebus
user’s manual
& Central Suite software installation CD
Setup Overview
The following diagrams depict overall views of how the Cerebus and the Digital Cerebus Systems are
integrated within a typical laboratory environment. The experiment subject area is both electrically and
optically isolated from the experiment operation and control area. In the Cerebus System, digitization and
amplification takes place inside the Front End Amplifier. The amplifier receives analog signals either
directly from electrodes or via analog headstages then performs differential amplification and digitization.
In the Digital Cerebus System, the neural signals are digitized by the CerePlex digital headstages before
they enter the Digital Hub. Both the Digital Hub and Front End Amplifier are also responsible for
converting the digitized signals into fiber optic data and transferring them to the processor via a fiber
optic link.