90
CMC?
Description:
Queries the measured/readback current output
value of the selected power supply
Query:
CMC?
Return
String:
1.234 A
The selected power supply has a measured output
current of 1.234 A.
CDVC?
Description:
Queries the display voltage and current of the
selected power supply.
Example:
CDVC?
Queries channel 1
Return
String:
100.23,1.567
Display voltage is 100.23 V and current is 1.567 A.
Format:
<voltage>,<current>
COUT
Description:
Configures/Queries the output state of the selected
power supply.
Example:
COUT ON
Sets the selected power supply’s output to ON.
COUT?
Queries the selected power supply’s output state.
COV
Description:
Configures/Queries the overvoltage protection
Summary of Contents for XLN10014
Page 90: ......
Page 94: ...Model XLN15010 XLN30052 XLN60026 High Power Programmable DC Power Supply USER MANUAL...
Page 113: ...17 voltage is within the specified range...
Page 204: ...108...
Page 206: ...110...