Measure
101
11.1.2 Time Measurements
Time measurements includes 11 types of time parameter measurements.
Figure 11.4
Time Measurements
1.
Period
: Period for every cycle in waveform at the 50% level, and positive slope.
2.
Frequency
: Frequency for every cycle in waveform at the 50% level ,and positive slope
3.
+ Width
: Width measured at 50% level and positive slope.
- Width
: Width measured at 50% level and negative slope.
4.
Rise Time
: Duration of rising edge from 10-90%.
5.
Fall Time
: Duration of falling edge from 90-10%.
6.
BWid
: Time from the first rising edge to the last falling edge, or the first falling edge to the last rising edge at the
50% crossing.
7.
+ Duty
: Ratio of positive width to period.
- Duty
: Ratio of negative width to period.
8.
Delay
Time from the trigger to the first transition at the 50% crossing.
9.
T@L
Time from trigger of each transition at a specific level and slope, include: Current, Max, Min, Mean, and Std-dev.
11.1.3 Delay Measurements
Delay measurements measure the time different between arbitrary two channels, including 10 types of delay measurements.
1.
Phase
: Calculate the phase difference between two edges.
2.
FRFR
: Time between the first rising edges of the two channels.
3.
FRFF
: Time from the first rising edge of channel A to the first falling edge of channel B.
4.
FFFR
: Time from the first falling edge of channel A to the first rising edge of channel B.
5.
FFFF
: Time from the first falling edge of channel A to the first falling edge of channel B.
6.
FRLR
: Time from the first rising edge of channel A to the last rising edge of channel B.
7.
FRLF
: Time from the first rising edge of channel A to the last falling edge of channel B.
8.
FFLR
: Time from the first falling edge of channel A to the last rising edge of channel B.
Summary of Contents for 2194