Serial Trigger and Decode
80
7.5 LIN Trigger and Serial Decode
7.5.1 Setup for LIN Signals
There are two steps of setting the LIN signal, connecting the signal to oscilloscope, specifying the parameters of each
input signal.
1. Touch
Decode
button to enter the
DECODE
function menu.
2. Touch
Decode
and select the desired slot (Decode1 or Decode2).
3. Touch
Protocol
and then select
LIN
by turning Universal Knob.
4. Touch
Signal
to enter the
SIGNAL
menu as shown in figure
.
Figure 7.28
CAN Signal Menu
5. Touch
Source
to select the channel that is connected to the LIN signal.
6. Touch
Threshold
and set the LIN signal’s threshold voltage level by
Universal Knob
. The threshold voltage level is
for decoding, and it will be regard as the trigger voltage level when set the trigger type to serial.
7. Touch
to return previous menu.
8. Touch
Configure
to enter the
BUS CONFIG
menu.
9. Touch
Baud
to set baud rate.
• The baud rate can be set as predefined value.
• If the desired baud rate is not listed, select
custom
option, press the Custom and turn the
Universal Knob
to
set the desired baud rate.
Summary of Contents for 2194