Trigger
49
Table
explains the relation between
Of Lines
,
Of Fields
,
Interlace
,
Trigger Line
and
Trigger Field
using an
Of Lines
value of 800.
Of Lines
Of Fields
Interlace
Trigger Line
Trigger Field
800
1
1:1
800
1
800
1, 2, 4 or 8
2:1
400
1, 12, 14, 18
800
1, 2, 4 or 8
4:1
200
1, 12, 14, 18
800
1, 2, 4 or 8
8:1
100
1, 12, 14, 18
Table 6.3
Parameters Relations
6. Press the
Sync
softkey to select
Any
or
Select
trigger mode.
•
Any
: Trigger on any of the horizontal sync pulses.
•
Select
: Trigger on the appointed line and field you have set.
– Press the
Line
or
Field
softkey.
– Turn the Universal Knob to set the value.
Table
list the line numbers per field for each video standard.
Standard
Field 1
Filed 2
NTSC
1 to 262
1 to 263
PAL
1 to 312
1 to 313
HDTV 720P/50, HDTV 720P/60
1 to 750
HDTV 1080P/50, HDTV 1080P/60
1 to 1125
HDTV 1080iP/50, HDTV 1080i/60
1 to 562
1 to 563
Table 6.4
Line Numbers Per Field
Triggering on a Specific Line of Video
Video triggering requires greater than 1/2 division of sync amplitude with any analog channel as the trigger source.
The example below is set to trigger on field 2, line 124 using the NTSC video standard.
1. Press the Setup key to enter the
Trigger
menu.
2. Press the
Type
softkey, then use the Universal Knob to select
Video
.
3. Press the
Source
softkey, then use the Universal Knob to select
CH 1
as the trigger source.
4. Press the
Standard
softkey, then use the Universal Knob to select
NTSC
as the trigger source.
5. Press the
Sync
softkey and set the option to
Select
; press the Line softkey and then turn the universal to set 22.
6. Press the
Line
softkey, then use the Universal Knob to select
022
as the trigger source.
7. Press the
Field
softkey, then use the Universal Knob to select
1
as the trigger source.
Summary of Contents for 2194
Page 1: ......
Page 82: ...Serial Trigger and Decode 82 Figure 7 29 LIN Trigger ...