Mechanical Parts
bkSpecto Ultrasound System
6-20
SM41406-4
Remove the GPU
Initial tasks in chronological order:
6.2.7 Remove the Main Mounting Plate
Proceduce:
1. Remove the two screws.
2. Gently lift and slide off the GPU.
Remove the On/Off Button
Initial tasks in chronological order:
6.2.7 Remove the Main Mounting Plate
Procedure:
1. Gently press out the button
M3X6