bkSpecto Ultrasound System
Troubleshooting
SM41406-4
4-83
The battery charging is handled can be ready by removing or connecting power cord. It will
be shown in a line like:
(ST:33->00)
Here are the states:
State
Comment
00
Battery is discharging (scanner running solely on
batteries)
11
Battery is charging (scanner is running on power cord
and battery is being recharged)
33
Standby (scanner is running on power cord and battery is
not being recharged)
Read Cycle count and serial number
1. Type
bdi;
and click on
Send
2. Read in Data output area the serial number
3. Read in Data output area the cycle count
All other data is for R&D
Configure a Scanner to Archive to a Network Drive (SW 6.6.4 and after)
The 1300 platform uses Windows 10 Embedded. All actions in this section are predicated on
that. Follow these steps to configure the scanner to copy exams on a network drive:
1. Conforming Microsoft Windows-based PC Status
2. Configure Microsoft Windows-based PC
3. Configure Network
4. Configure Scanner
Conforming Microsoft Windows-based PC Status
NOTE:
This is not to be done on the scanner.
Before proceeding, determine whether the PC is member of a domain or a standalone PC.
To Determine PC Status – Domain or Standalone:
fc
Full electrical charge measure in
mAh
To extend the battery life, we never charge the batteries to full
electrical charge, so rc wil never be equal to fc
Default value is approx. 3800 mAh
V
Measured voltage in mV
Default value is approx. 24,49 V
I
Measured current in mA
A negative value: battery being drained (4 digit value)
A positive value: battery being charged (4 digit value)
(The 2 digit value in the example above is standby current )
fl
Used by R&D only
t
Temperature in degrees Celsius Default value is approx. 26 degrees Celsius (?)