bkSpecto Ultrasound System
Mechanical Parts
SM41406-4
6-9
3. Unfasten the 8 screws and remove the chassis.
Note:
Remember to insert the power button extender when assembling the scanner again.
Remove Battery Assembly
Initial tasks in chronical order:
6.2.2
Remove the Left Chassis
Procedure:
1. Check all the fan wires are labelled with the fan number. If a label is missing add a new
label.
2. Disconnect the six fan wires from the callisto main board. Labels on the wires indicate
the fan number.
Caution:
Remember to insert the wires in the order as shown above when assembling the scanner. Fan
numbers and locations are listed in Table 6-3.
M4X1
M3X6
Wires from 4 & 5
Wires from 0,1,2 & 6
Metal
Brackets
Summary of Contents for bkSpecto 1300