bkSpecto Ultrasound System
Troubleshooting
SM41406-4
4-35
6. Is the connection not a success make sure the information from the
DICOM Network
Description
questionnaire has been added correctly.
7. Select
OK
twice.
PACS Setup
Note:
Scanner AE Title
must be present before testing can be done.
Setting up PACS includes choosing parameters for
Archiving Method
,
Auto Send
Timeout
, and
Storage Commitment
. Table 4-12, Table 4-13, and Table 4-14 contain the
possibilities and definitions.
Method
Definition
Manual
Images are sent manually to PACS.
Auto
Tap the touch screen
End Exam
button. The current examination
will end and all images and clips (except 3D) will be sent. Starting
a new exam will also auto-archive images and clips for the
previous patient. This is the default
Archiving Method
.
Table 4-12: Available archiving method options for PACS.
Method
Definition
Send complete examination at end of
examination
Scanner will try to send after end of examination.
Immediately
Scanner will try to send immediately.
5 min, 10 min, 15 min, 30 min, 60 min
Scanner will try to send after the specified time.
Table 4-13: Available auto send timeout options for PACS.
Method
Definition
Not Used
Storage Commitment
feature is disabled.
Use PACS Server
The PACS Server will handle the
Storage
Commitment
request.
User Storage Commitment Server You can specify a
Storage Commitment Server
using
the following fields:
Name
,
AE Title
,
IP Address
and
Port Number
.
Note:
Once accepted,
Name
cannot be edited later.
Table 4-14: Available storage commitment options for PACS.