Troubleshooting
bkSpecto Ultrasound System
4-34
SM41406-4
Search Filter Values
Definition
This System Only
The AE title of the scanner will then be used to differentiate.
Ultrasound System Only
(0008,0060)-tag must be ultrasound (US) in the record in
the worklist.
All Systems
No filtering. All modalities on the worklist will appear on the
scanner.
Table 4-11: Search filter possibilities when setting up a worklist.
To Configure the DICOM Worklist:
1. Access the DICOM Setting Dialog according to 4.29.1.
2. Select
Configure
(next to
Worklist
field).
3. Enter
Name
,
AE Title
,
IP Address
, and
Port Number
.
Note:
Once accepted,
Name
cannot be edited.
4. Select the desired
Search Filter
value from the drop-down menu.
5. Select
Test
.