Limitations
- Store RIGHTEST Blood Glucose Test Strips in the original capped vial at temperatures between 4°C to
30°C (39°F to 86°F) and relative humidity below 90%. Do not freeze.
- RIGHTEST Blood Glucose Test Strips are designed for using with capillary whole blood samples. Do
not use serum or plasma samples.
- Inaccurate test results may be obtained at high altitude more than 3,048 meters (10,000 feet) above
sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- RIGHTEST Blood Glucose Monitoring System has not been validated for use on neonates.
- Test result may be interfered in the presence of abnormally high concentration of interferences or
HCT, please refer to strip insert for the detail.
Sampling Error
Do not apply the blood drop to the sample entry of the strip before RIGHTEST
Blood Glucose Meter displays " ". If you do so, the meter will display " "
and " " accompanied by beeps (if volume is turned on). Please discard the test
strip and repeat the test with a new RIGHTEST Blood Glucose Test Strip.
Meter Malfunction
If RIGHTEST Blood Glucose Meter will not turn on, please follow the steps below:
1. Open the battery cover and remove the battery.
2. Wait for 5 minutes and reinsert the battery as described " Meter activation and Battery change " on
page 16 - 17.
RIGHTEST Blood Glucose Meter should work normally after finishing the above steps. If not, please
contact customer service.
Error messages and Trouble shooting