31
30
Performing a Test in Order
View Window Appearance
Make sure your blood sample covers the whole
area of the view window to get an accurate test
result. An insufficient blood sample will result in
an error message ("Er4"). If this occurs, repeat
the test with a new test strip.
CAUTION
- Check the expiration date printed on the package every time you use the strip. Do not use
expired RIGHTEST Blood Glucose Test Strips.
- Use each RIGHTEST Blood Glucose Test Strip immediately after taking it from the vial.
- Do not reuse RIGHTEST Blood Glucose Test Strips.
- Wait at least 30 minutes to test after entering another location with a different ambient
temperature.
Insufficient blood sample
Enough blood sample
CAUTION
- Test results may vary if blood samples are taken from different sites or under certain
conditions where glucose levels can change such as: following a drink, a meal, an insulin
dose or exercise. In these cases, only the fingertip should be used.
- DO NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood
glucose).
- Fingertip samples can show rapid changes in glucose faster than palm or forearm samples.
- Consult your healthcare professionals before sampling from your palm or forearm.
- As the blood flow taken from forearm is slower than fingertip or palm, we recommend using
a special lancing device with a Clear Cap for testing sites other than fingertip.
- If you use the normal lancing device for your palm or forearm, the blood sample may not be
enough for the meter.
- Periodically compare the test system to a laboratory test system that is known to be well
maintained and monitored by a healthcare provider.
Summary of Contents for Rightest GT100