BIONIME Rightest GT100 User Manual Download Page 19

34

35

Quality Control Test

CAUTION

 

- Each time you open a new bottle of RIGHTEST Control 

Solution, write the expiration date on the label. RIGHTEST 
Control Solution is good for 3 months after opening the 
bottle, or until the expiration date printed on the label, 
whichever comes first.

Possible reasons your Control Solution results are out of range :
- Your RIGHTEST Control Solution is expired or has been opened more than 3 months ago.

 

- Your RIGHTEST Blood Glucose Test Strip has expired.

  

- You left the cap of RIGHTEST Blood Glucose Test Strip vial or the GC550 Control Solution off for a 

period  of time.

- You didn't perform the test procedure correctly. 
- Malfunction of RIGHTEST Blood Glucose Meter or Test Strip.
If control solution test results are out of range, your RIGHTEST Blood Glucose Monitoring System 
may not be working properly. Repeat the quality control test. If your control solution results outside 
the range still exist, do not use RIGHTEST Blood Glucose Monitoring System to test your blood 
glucose and contact Bionime Customer service.

Quality Control Test

Use RIGHTEST Control Solutions when testing your RIGHTEST GT100 Blood Glucose Monitoring 
System under the Control Solution Mode. If the test result is within the Control Solution Range printed 
on the strip vial label, RIGHTEST Blood Glucose Monitoring System passes the Quality Control Test. 
That means your RIGHTEST Blood Glucose Monitoring System is working correctly.

Control Solution Range:

Example of Control Solution Range 
printed on your test strip vial label.

When should a Quality Control Test be performed ?

 

- To ensure that your RIGHTEST Blood Glucose Meter and Test Strip are working together properly.
- To confirm that you are following the correct procedure.
- To prepare for your initial blood glucose test.

 

- To check RIGHTEST Blood Glucose Test Strip when you open a new vial of strips.

 

- To check your RIGHTEST Blood Glucose Meter after it has been dropped, damaged or exposed to 

liquids.

- When you suspect that your test results are inaccurate, or if your test results are not consistent with 

the way you feel.

Example of Expiration Date:

NOTE

 

- If you want to purchase new RIGHTEST Control Solutions, please contact  customer service.

Level 1
Level 2
Level 4

Control Solution Range     mg/dL

Normal:                      83 - 113 mg /dL 

High:                        237 - 321 mg/dL

Exp. Date 2019-12-31

mmol/L

4.6 - 6.3 mmol/L

13.2 - 17.8 mmol/L

Summary of Contents for Rightest GT100

Page 1: ...g City 40242 Taiwan Tel 886 4 2369 2388 Fax 886 4 2261 7586 http www bionime com E mail info bionime com Made in Taiwan User s Manual GT100 Rev Date 2019 10 Blood Glucose Monitoring System BIONIME Gmb...

Page 2: ......

Page 3: ...ofessional use Designed for in vitro diagnostic use only testing requires just a small amount of fresh capillary whole blood from either the fingertip palm or forearm If you have any questions regardi...

Page 4: ...s RIGHTEST Blood Glucose Monitoring System has not been validated for use on neonates Therefore it s not intended for use of neonates RIGHTEST Blood Glucose Monitoring System has not been used for art...

Page 5: ...bout RIGHTEST System The RIGHTEST Blood Glucose Monitoring System The RIGHTEST Blood Glucose Meter The RIGHTEST Blood Glucose Test Strip Before Testing Meter activation and Battery change Setting Up Y...

Page 6: ...ip Package Insert 4 RIGHTEST GC550 Control Solution Package Insert 5 RIGHTEST GT100 Blood Glucose Meter with one CR2032 battery installed 6 RIGHTEST GT100 Blood Glucose Test Strip 10 25 pcs 7 RIGHTEST...

Page 7: ...g the meter please check details in following sections Display Window Shows your test result messages and relative information Test Strip Port Insert test strip here until you hear a click Battery Cov...

Page 8: ...the battery is low or must be replaced Indicates when to apply the blood sample Test strip Current date under time mode or testing date under memory mode Indicates if the environmental temperature li...

Page 9: ...Insert If the RIGHTEST Blood Glucose Meter and Test Strips are exposed to a high temperature difference please wait 30 minutes before measurement View Window This window is yellow before applying blo...

Page 10: ...s the main button or insert a strip to activate your meter NOTE After pressing the main button for 3 seconds screen display will turn off Keep pressing the main button to enter the setting data Contin...

Page 11: ...m Once the day is confirmed the Time setting will appear 1 2 3 4a 4b 5 6 7a 7b 4 Time format 12 24H selection With the time format blinking press the Up or Down button to switch between 12H and 24H Pr...

Page 12: ...return to time screen 1 How to turn on RIGHTEST Blood Glucose Meter Press the Main button or insert a test strip 2 Manual Power off If you want to turn off RIGHTEST Blood Glucose Meter press and hold...

Page 13: ...oving RIGHTEST Blood Glucose Test Strip 1 Hold RIGHTEST Blood Glucose Test Strip as shown in insertion instructions 2 Rotate RIGHTEST Blood Glucose Test Strip counterclockwise and pull up simultaneous...

Page 14: ...r Hub Lancet carrier Release button Safe switch Depth adjustable cap Performing a Blood Test 2 Pull the cap and hub off in opposite directions remove the cap 3 Insert a new disposable lancet firmly in...

Page 15: ...ased on skin type for soft or thin skin for average skin for thick or calloused skin 8 NOTE RIGHTEST Blood Glucose Meter will automatically detect the Code number on the strip You don t have to check...

Page 16: ...T Blood Glucose Test Strip Record the opening date of a newRIGHTEST Blood Glucose Test Strip vial Discard the vial of test strips 12 months after opening Performing a Blood Test Alternative site testi...

Page 17: ...sults may vary if blood samples are taken from different sites or under certain conditions where glucose levels can change such as following a drink a meal an insulin dose or exercise In these cases o...

Page 18: ...f you are experiencing symptoms that are not consistent with your blood glucose test results and you have made sure to follow all instructions in this manual contact your healthcare professional immed...

Page 19: ...EST Control Solutions when testing your RIGHTEST GT100 Blood Glucose Monitoring System under the Control Solution Mode If the test result is within the Control Solution Range printed on the strip vial...

Page 20: ...for average reading but can still be recalled The Control Solution Test result will be shown with symbol on the screen Control Solution Test should be conducted between 15 40 C 59 104 F Before and ap...

Page 21: ...ur blood glucose values You will see the number of the calculated days in the lower left corner and the number of the calculated readings in the lower right corner 3 Non Averaging set and canceled You...

Page 22: ...mory screen CAUTION You have to set the time and date to activate the average function The and results the Control Solution results and test results made out of normal temperature range 10 C 50 F 40 C...

Page 23: ...eliminate non visible pathogens The following disinfecting wipe has been tested and may be used to clean and disinfect the meter and lancing device CAVIWIPES DISINFECTING TOWELETTES manufacturer Metre...

Page 24: ...e of the meter 5 years Maintain the Products CAUTION Users should wash their hands thoroughly with soap and water after handling the meter lancing device or test strips Please examine your LCD screen...

Page 25: ...r support This error message can also indicate that you may have used the wrong test strips Please check the test strip vial to ensure you are using RIGHTEST Test Strips 47 46 Error messages and Troub...

Page 26: ...EST Blood Glucose Meter will blink the symbol with the thermograph symbol displaying on screen 1a 3 37 CAUTION A blood sample should ONLY be applied to the test strip after the test strip has been ins...

Page 27: ...esence of abnormally high concentration of interferences or HCT please refer to strip insert for the detail Sampling Error Do not apply the blood drop to the sample entry of the strip before RIGHTEST...

Page 28: ...utomatically 2 minutes after last user action or press the button for 3 seconds Measurement Technology Measurement Calibration Measuring Range Test Time 10 600 mg dL 0 6 33 3 mmol L 5 seconds Oxidase...

Page 29: ...f used symbols EC Representive CE mark with No of notified body Consult the instruction for use Warranty Bionime Corporation warrants this product will be free from defects in materials and workmanshi...

Page 30: ...No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan Product complied with Medical Device Directive 93 42 EEC EC Rep Bionime GmbH Tramstrasse 16 9442 Berneck Switzerland E mail info bionime ch...

Reviews: