Temperature Error
In order to get an accurate test result, perform testing between 10 - 40°C (50 - 104°F).
1. When the ambient temperature is 0 - 9°C (32 - 48°F)
or 41 - 50°C (106 - 122°F)
the " " warning symbol will blink (1a). You can still do the test but the test
result is for reference. Tests in the memory screen this value will be flagged with
" " symbol (1b). Repeat the test in an area with temperature ranges between
operating range
(10 - 40°C or 50 - 104°F).
2. When you move from an area with temperature outside the operating range to
another area with temperature inside operating range, please wait for 30 minutes
before you test.
3. When the temperature is below 0°C (32°F) or over 50°C (122°F), RIGHTEST
Blood Glucose Meter will blink the " " symbol with the thermograph symbol
displaying on screen.
1a
3
37
CAUTION
A blood sample should ONLY be applied to the test strip after the test strip has been inserted
correctly and the Meter's screen is showing the image of the test strip and a blood drop
flashing symbol. If the blood drop flashing symbol is NOT showing on the screen, do not
apply a sample to the test strip. Please re-insert the unused test strip correctly. The screen
MUST show the image of the test strip accompanied by the blood drop symbol flashing (this
symbol take 3 seconds to show after correctly inserting the test strip) in order for a sample to
be applied. Please consult your User's Manual and/or contact Customer Service for support
on how to correctly insert a RIGHTEST Test strip.
1b
Error messages and Trouble shooting
Error messages and Trouble shooting