44
Drain Waste & Prime Buffers
Priming the instrument should be done daily before the first run of the day. The user should prime until there are no remaining bubbles
in the carboy tubing or probe coil. When priming the manifold, prime until the stream coming from the port is steady and even across
the 12 ports. Priming should also be done after refilling the carboys, this does not require priming the manifold. This function is found
under Tools-> Instrument Operations-> Drain Waste & Prime Buffers
Prime Probe
• Select Prime from the Select Operation section.
• When Prime is selected, the following are enabled: Probe, Buffers, DePar and Trough.
• To prime probe, select Probe and adjust the duration of prime in seconds as needed, then click Prime Solution.
• When priming is complete, click Close or select another priming option.
Prime Buffers
• To prime, select Prime under Select Operation, then click Buffers under Select a Priming Option.
• Select Aqua Rinse or Buffer
• Adjust the duration of prime in seconds as needed and click Prime Solution.
• When priming is complete, click Close or select another priming option.