![BioCare Medical VALENT Manual Download Page 23](http://html.mh-extra.com/html/biocare-medical/valent/valent_manual_2753390023.webp)
15
Home Screen Features
Name
Function
1.
Menu Bar
Select from the menu bar to navigate instrumental functions and access information.
2.
Run Designation Buttons
Allows user to Prepare Slides, Map Slides, Print Reports or Cancel Run.
3.
Run End Time
Displays the run end time.
4.
Reagent Rack
The reagent rack is a stainless steel rack with 44 positions for 20mL vials that can easily slide
in and out of the instrument. The reagent rack remains below room temperature throughout a
run. Use the mouse cursor to hover over a reagent vial to view reagent information during a run.
5.
System Lid
Displays whether system lid is opened or closed.
6.
Reagent Rack Lock/Unlock
Status
Displays whether the reagent rack is locked or unlocked.
7.
Reagent Rack Temperature
Displays the temperature of the reagent rack to ensure reagents are stored at an acceptable
temperature range.
8.
Carboy Status
The carboy status indicates if there is sufficient DePar, Aqua Rinse, Antigen Retrieval (AR)
and Wash Buffer available in the bulk reagent carboys. The carboy status also displays the
amount of non-hazardous waste and hazardous waste in the carboy.
A reminder will be displayed when bulk reagent carboys are low and when waste carboys need
to be emptied.
9.
Slide Tray Position
The VALENT Software displays the location of each slide on the slide tray numbered from
1-12. Hovering over a mapped slide allows the user to view slide information and protocol
during a run.
10.
Slide Tray Lock/Unlock
Status
Displays whether a slide tray is locked or unlocked. Slide trays are locked during a run.
11.
Slide Tray Status
The VALENT Software displays the slide trays and the slide positions graphically on the user
interface. The slide trays are designated A-D. Tray status indicates whether the tray is in
process (red) or available (green).